This method of testing can be used to evaluate electrodes and amplifiers for damage. This is recommended if receiving inconsistent or inaccurate data during a session.
Testing Impedance Circuitry
Without electrodes connected, this should report the following values depending on the amplifier used:
A202/A404: 50k; one value for each Signal and Reference
U-Wiz/Q-Wiz: -2/256k, 256k, 0k; one value for each Channel
Optima 2/Optima+ 4: 125k; one value for each Signal and Reference
J202/J404: 2000k; one value for each Signal and Reference
This value indicates an ‘open’ connection, meaning the electrodes are not being touched together or on a client’s scalp. If this value is not reported, it indicates damage to the amplifier.
Insert electrodes into the Signal A, Reference A, and Ground ports.
If the amplifier reports an open connection with no electrodes connected, and drops down to 1k when electrodes are touched together, this tells us that both the electrodes and the amplifier's impedance circuitry are functioning correctly.
Testing the Signal Response
Press Ctrl+F7 to return to the Therapist/Wave Screen.
If the electrodes are still connected to the amplifier, we may observe some ambient noise on the RAW traces. Touch the electrodes again and see if this ambient noise flatlines. If it does not, try to move away any nearby electrical devices (smartphones, smartwatches, etc.) and see if the signal improves.
If the noise does not flatline, or is severe, that may indicate damage to the amplifier. Additionally, if the 'center' of the raw trace is above or below the middle line, or the signal appears to be 'wandering' up and down, that indicates damage to the amplifier.
Observe the Spectral FFT Bars at the bottom of the Therapist/Wave Screen - identify if there is a large and consistent spike in the signal and at what frequency it is. This can sometimes indicate damage to the amplifier or outside interference.